| Specification |
SR100 |
| Electric stirring motor |
220–240
V, 50/60 Hz, single phase, 200 W, 1/4 HP Maximum rotation speed: 450 rpm (±8%) Starting torque: 3.0 N·m Rated torque: 1.54 N·m Built-in gearbox ratio: 3:1 to 180:1 Controller: Delta, UL listed |
| Pneumatic stirring motor |
Power:
1/4 HP Maximum rotation speed: 450 rpm (±8%) Compressed air pressure: 30–80 psi (0.2–0.55
MPa) Working temperature: −10°C to +70°C For continuous operation: set speed to 80% of
maximum rpm |
| Pneumatic motor lubrication |
Motor
lubricant: Daphne Eponex EP-No. 1 Check, replenish or replace every 500 hours |
| Air lubrication |
Type:
JIS K2213-1 (ISO VG32) Recommended products: Shell Vitrea Oil 32, JOMO
Turbine 32, COSMO Turbine 32 Lubrication rate: 2–3 drops per minute |
| Pneumatic motor operating precautions |
Disconnect
the compressed air supply when not in use, unattended or overnight. Do not run the motor unloaded. Set the air adjuster to minimum for the initial
run. Check that the motor operates correctly before increasing airflow. Stop the motor and disconnect the air supply
before inspection. Do not inspect while the motor is running. Keep the exhaust port clear of debris and
blockages. Check the air lubricant and drain water from the
air filter every 2–3 days. |
| Port configuration |
Lid:
150 mm tri-clamp (nominal 6″) Center motor port: 75 mm tri-clamp (nominal 3″) Drain port: 150 mm tri-clamp (nominal 6″) Drain port on/off valve: 50
mm tri-clamp (nominal 2″) Vacuum port: 40 mm tri-clamp (nominal 1.5″) Condenser recirculating fluid port: 15 mm female NPT (nominal 1/2″) Vessel jacket recirculating fluid ports: 25 mm female NPT (nominal 1″) |
| 50 L / 100 L collection tank — standard lid |
150
mm tri-clamp lid (nominal 6″), with 4 × 40 mm tri-clamp ports (nominal 1.5″) |
| 50 L / 100 L collection tank — additional lid |
150
mm tri-clamp lid (nominal 6″), with: 2 × 15 mm FNPT ports with dip tubes (nominal 1/2″) 1 × 15 mm FNPT port (nominal 1/2″) 1 × 5 mm FNPT port (nominal 1/4″) 1 × 50 mm tri-clamp port (nominal 2″) |
| 50 L / 100 L collection tank — other ports |
Top
of tank: 2 × 40 mm tri-clamp
ports with sight glasses (nominal 1.5″) Drain port: 40 mm tri-clamp (nominal 1.5″) |
| Main vessel |
Volume:
100 L Material: 316L stainless steel Interior finish: Electropolished |
| Agitator |
Structure:
Three-blade, dual-layer propeller (standard) Style: Quick-change Material: 316L stainless steel Detachable length: 510 mm |
| Center fluid jacket |
Volume:
22 L Material: 304 stainless steel |
| Outer insulation jacket |
Sealed Material: 304 stainless steel |
| Collection tank |
Option
1: 2 × 10 L, 316L stainless steel, with manifold Option 2: 1 × 50 L, 304 stainless steel, with
dolly; 395 mm diameter × 595 mm height; 760 mm height
with dolly Option 3: 1 × 100 L, 304 stainless steel, with
dolly; 395 mm diameter × 1,090 mm height; 1,255 mm height
with dolly |
| Material |
All
wetted parts on the main reactor body: 316L stainless steel Remaining reactor components: 304 stainless
steel |
| Main condenser |
Style:
Shell-and-tube heat exchanger Number of tubes: 70 Condensing area: 0.89 m² |
| Pressure |
Ultimate
vacuum: 1 Torr |
| Temperature range |
−55°C
to +250°C (−67°F to +482°F) |
| Temperature sensor/display |
Standard:
BOOST.TTA.0101, −200°C to +480°C, 50.5 clamp Explosion-proof: BOOST.TTA.0102, −200°C to
+480°C, 50.5 clamp |
| Sealing gasket |
Food-grade
Viton |
| Main mounting frame |
Heavy-duty
50 × 50 mm 304 stainless steel
frame, mounted on four casters |
| Drain port ground clearance |
Drain
port: 495 mm Drain valve: 355 mm |
| Unit / shipping weight |
116
kg / 159 kg |
| Unit / shipping dimensions, W × D × H |
970
× 810 × 2,080 mm |
| Compliance |
ETL
tested to UL and CSA standards C1D1 peer review by PSI (optional) |